On the characterization of an AHF cavity radiometer and its traceability to WRR/SI

dc.contributor.authorBalenzategui Manzanares, J. L.es
dc.contributor.authorDe Lucas Veguillas, Javieres
dc.contributor.authorCuenca, J.es
dc.contributor.authorMolero, M.es
dc.contributor.authorRomero, M. C.es
dc.contributor.authorFabrero, F.es
dc.contributor.authorSilva, J. P.es
dc.contributor.authorMejuto, E.es
dc.contributor.authorIbañez, F.J.es
dc.contributor.funderMinisterio de Industria, Economía y Competitividades
dc.date.accessioned2023-02-20T08:45:57Z
dc.date.available2023-02-20T08:45:57Z
dc.date.issued2022-02-10
dc.description.abstractIn a complementary way to the comparison to WSG to get traceability to WRR (and consequently, to SI), a solar-type cavity radiometer can also be characterized, determining the deviations of the instrument from the ideal realization of the principle of electrical substitution and obtaining its total measurement uncertainty. This work summarizes different techniques and procedures applied for the characterization of an Eppley AHF radiometer. The approach for characterization is based on the analysis of the measurement model function of the instrument. Some results obtained from calibration and testing (voltmeter, area of the precision aperture, resistance of the leads, non equivalence factor), and from numerical simulation (effective absorptance, scattering) are presented. According to these results, current value of standard uncertainty for this instrument is about 0.28% but it is expected that further improvements in the equipment and tests can reduce this figure below 0.1% (1000 ppm) in the near future.es
dc.description.peerreviewedPreprintes
dc.description.sponsorshipMinisterio de Industria, Economía y Competitividades
dc.identifier.citationProceedings of the Symposium on Radiation Measurement (held in conjunction with IPC-XIII, FRC-V and IPgC-III)es
dc.identifier.otherhttp://documenta.ciemat.es/handle/123456789/1436es
dc.identifier.urihttp://hdl.handle.net/20.500.12666/836
dc.language.isoenges
dc.publisherCIEMATes
dc.relationProject “Desarrollo de Patrones Primarios de Irradiancia Solar Basados en Radiómetros Absolutos de Cavidad (DEPRISACR)”, Ref. CGL2017-87299-Pes
dc.rights.accessRightsinfo:eu-repo/semantics/openAccesses
dc.subjectSolar irradiancees
dc.subjectCalibrationes
dc.subjectRadiometryes
dc.subjectCharacterizationes
dc.subjectUnvertaintyes
dc.subjectTraceabilityes
dc.titleOn the characterization of an AHF cavity radiometer and its traceability to WRR/SIes
dc.typeinfo:eu-repo/semantics/conferenceObjectes
dc.type.coarhttp://purl.org/coar/resource_type/c_8544es
dc.type.hasVersioninfo:eu-repo/semantics/publishedVersiones

Archivos

Bloque original

Mostrando 1 - 1 de 1
Cargando...
Miniatura
Nombre:
Balenzategui et al - IPC-XIII Symposium.Open.2.pdf
Tamaño:
1.26 MB
Formato:
Adobe Portable Document Format
Descripción:
On the characterization of an AHF cavity radiometer and its traceability to WRR/SI

Bloque de licencias

Mostrando 1 - 1 de 1
No hay miniatura disponible
Nombre:
license.txt
Tamaño:
4.82 KB
Formato:
Item-specific license agreed upon to submission
Descripción: